Next: Once Per Data Run
Up: No Title
Previous: No Title
- Record the shiftchange in the general logbook and keep this logbook up to date.
Record all unusual events (ie HV trips etc.) in this logbook.
- Run XCALMON, the monitoring software by pressing the button on
atlhp-h6 or entering
xcalmon at the prompt on atlhp-h6 (ensure
the
DAQ software and purity monitoring software are not running concurrently).
Print a summary (press the 'print monitor histo' button) and append to
XCALMON folder. Percent differences should be within 2%. Pedestals within
1 ADC count. If they are not, call the expert.
- Run purity monitoring software by entering
bm_run at the
prompt on atlhp-h6 (ensure the DAQ software and xcalmon are not running
concurrently). Compare purity spectra to reference plots. Print and append
to purity log.
- Fill in the checklist and append to Check-list binder.
- Monitor diskspace by entering
bdf at the unix prompt on atlhp-h6 (or by looking at the graphical
disk space monitoring window, if it is activated). If /data2 disk becomes
full, call expert.
- Stability runs should be taken once per day using 120GeV electrons on impact points H and I.
Matthew Adam Dobbs